- 7″ Color LCD with touchscreen
- HDMI for external video output
- Save up to 5,000 cases at Spot mode
- USB memory storage in CSV file to EMR
- Smart, clear display
- Li-Ion battery up to 4 hrs
Parameters | Product Specifications | Value |
---|---|---|
Hardware specification | Dimension (W x H x D) | 188 x 180 x 62 mm |
Weight | Approx. 1.5kg | |
Display | 7” color TFT-LCD full touch screen | |
Resolution | 800 x 480 pixels | |
Parameters | SpO2, Pulse Rate, Systolic BP, Diastolic BP, Mean BP, Temperature(IR), EtCO2, FiCO2, Airway Respiration Rate | |
Trace | 2 waveforms : SpO2, EtCO2 Sweep speed : 6.25, 12.5, 25, 50 mm/sec | |
Indicators | Categorized alarms (3 priority levels) SpO2 pitch tone, Battery status, External power LED | |
Interfaces | DC input connector : 15VDC, 2.0A LAN digital output for transferring data Nurse call system connection: • 0.3A at 125VAC • 1A at 24VDC External monitor connection: HDMI USB barcode reader USB data storage | |
Battery | Rechargeable Li-ion battery 4 hours for continuous operation (max.) | |
Data Storage | 168 hours trends, 5000 cases of patient data | |
Language | English, French, Spanish, Italian, Germany, Chinese, Russian, Czech, Bulgarian, Polish, Portuguese, Romanian, Hungarian, Turkish, Korean | |
Environment | Temperature | Operation: 5 ~ +40 °C (41 ~ 104 °F) Storage: –20 ~ +60 °C (–4 ~ +140 °F) |
Humidity | Operation: 30% ~ 85%, Storage: 10% ~ 95% (Package ) | |
Altitude | Operation: 525 ~ 795 mmHg (70 ~ 106 kPa) Storage: 375 ~ 795 mmHg (50 ~ 106 kPa) | |
SpO2 | Saturation range | 0 to 100% |
Saturation accuracy | 70 to 100% ±2 digits 0 to 69% unspecified | |
Pulse rate range | 30 to 254 bpm | |
Pulse rate accuracy | ±2 bpm | |
NIBP | Method | Oscillometric |
Operation Mode | Manual/Automatic/Continuous | |
Measurement range | Systolic : 40 to 260 mmHg MAP : 26 to 220 mmHg Diastolic : 20 to 200 mmHg | |
Accuracy | mean error: less than ±5 mmHg standard deviation : less than 8 mmHg Meets accuracy requirements of ANSI/AAMI SP10:1992 and 2002 (*SunTech) | |
Temperature | Measurement range | 34 to 42.5℃ (93.2 to 108.5℉) |
Accuracy | 36 to 39℃ : ±0.2℃ Other: ±0.3℃ | |
Sidestream CO2 (Optional) | Measurement range | 0 to 150 mmHg, 0 to 19% |
Accuracy | 0-40mmHg ±2 mmHg, 41-70mmHg ±5% of reading 71-100mmHg ±8% of reading, 101-150mmHg ±10% of reading | |
Respiration rate | 2 to 150 breaths per minute | |
Respiration accuracy | ±1 breath per minute | |
Measurement range | 0 to 150 mmHg, 0 to 19% | |
Accuracy | 0-40mmHg ±2 mmHg, 41-70mmHg ±5% of reading 71-100mmHg ±8% of reading, 101-150mmHg ±10% of reading | |
Respiration rate | 0 to 150 breaths per minute | |
Respiration accuracy | ±1 breath per minute |